a
|
|
HI 9024C
|
HI 9025C
|
|
Range:
|
pH
|
0.00 to 14.00
|
0.00 to 14.00
|
|
mV(ISE)
|
------
|
+/-399.9
|
|
mV(ORP)
|
------
|
+/-1999
|
|
degC
|
0.0 to 100.00
|
0.0 to 100.0
|
|
|
|
|
|
Resolution:
|
pH
|
0.01
|
0.01
|
|
MV(ISE)
|
------
|
0.1
|
|
MV(ORP)
|
------
|
1
|
|
degC
|
0.1
|
0.1
|
|
|
|
|
|
Accuracy(@20 degC / 68 degF):
|
pH
|
+/-0.01
|
+/-0.01
|
|
MV(ISE)
|
------
|
+/-0.2
|
|
MV(ORP)
|
------
|
+/-1
|
|
degC
|
+/-0.5
|
+/-0.5
|
|
|
|
|
|
Typical EMC Deviation:
|
pH
|
+/-0.01
|
+/-0.01
|
|
mV(ISE)
|
------
|
+/-0.3
|
|
mV(ORP)
|
------
|
+/-1
|
|
degC
|
+/-0.1
|
+/-0.1
|
|
|
|
|
pH Calibration:
|
|
Automatic, 5 selectable standard buffers
(pH 4.01, 6.86, 7.01, 9.18 or 10.01)
|
|
Offset Calibration:
|
|
+/-1 pH
|
|
Slope Calibration:
|
|
From 70 to 108%
|
|
Temp.Compensation:
|
|
Automatic or Manual from 32 to 212 degF (0 to
100 degC)
|
|
|
|
|
|
Electrode:
|
|
HI 1230B combination gel-filled double-junction
with BNC & 1 m (3.3') cable (included)
|
|
Temperature Probe:
|
|
HI 7669/2W with 1 m (3.3') screened cable (included)
|
| |
|
|
|
Input Impedance:
|
|
1012ohm
|
|
Battery:
|
|
4 x 1.5 volt AA
|
|
Battery Life:
|
|
400 hours of continuous use
|
|
|
|
|
|
Environment:
|
|
32 to 122 degF (0 to 50 degC); RH 100%
|
|
Dimensions:
|
Meter
|
196 x 80 x 60 mm (7.7 x 3.1 x 2.4")
|
|
Kit
|
340 x 230 x 80 mm (13.8 x 9 x 3.1")
|
|
Weight:
|
|
425 g (15 oz.); Kit: 1.3 Kg (3.0 lb.)
|